Biological evaluation of medical devices - Part 13: Identification and quantification of degradation products from polymeric medical devices

Abstract

ISO 10993‑13:2010 provides general requirements for the design of tests in a simulated environment for identifying and quantifying degradation products from finished polymeric medical devices ready for clinical use.

ISO 10993‑13:2010 describes two test methods to generate degradation products, an accelerated degradation test as a screening method and a real-time degradation test in a simulated environment. For materials that are intended to polymerize in situ, the set or cured polymer is used for testing. The data generated are used in the biological evaluation of the polymer. ISO 10993‑13:2010 considers only non-resorbable polymers. Similar, but appropriately modified procedures may be applicable for resorbable polymers.

ISO 10993‑13:2010 considers only those degradation products generated by a chemical alteration of the finished polymeric device. It is not applicable to degradation of the device induced during its intended use by mechanical stress, wear or electromagnetic radiation or biological factors such as enzymes, other proteins and cellular activity.

Begin

2024-11-04

WI

00206112

Planned document number

DIN EN ISO 10993-13 rev

Project number

02703456

Responsible national committee

NA 027-07-12 AA - Biological evaluation of medical devices  

Responsible european committee

CEN/TC 206 - Biological and clinical evaluation of medical devices  

Responsible international committee

ISO/TC 194/WG 2 - Degradation aspects related to biological testing  

previous edition(s)

Biological evaluation of medical devices - Part 13: Identification and quantification of degradation products from polymeric medical devices (ISO 10993-13:1998); German version EN ISO 10993-13:2009
2009-08

Order from DIN Media

Contact

Dr.

Susann Minkwitz

Am DIN-Platz, Burggrafenstr. 6
10787 Berlin

Tel.: +49 30 2601-2765
Fax: +49 30 2601-42765

Send message to contact